OCCUPATIONAL ALLERGIES AND ASTHMA: ACOEM GUIDELINE:
SPIROMETRY IN THE OCCUPATIONAL SETTING
Abstract
Improved quality and standardization of spirometry testing and interpretation
of results are critically important in the occupational setting. This position
paper hopes to contribute to that goal by increasing the occupational medical
community's awareness of the importance and complexities of spirometry testing.
The position paper reviews: basic principles of spirometry and indications for
spirometry in occupational medicine; essential criteria for assuring validity of
spirometric results; and proper interpretation of results, including selection
and race-adjustment of predicted values, comparison with predicted values, and
assessment of loss of function over time, response to a bronchodilator, and
acute changes associated with workplace exposures. ACOEM makes detailed
recommendations in each of these areas and key points are summarized in tables
throughout the statement.
Table of Contents
Spirometry In The Occupational Setting
Principles Of Spirometry
Indications For Spirometry In Occupational
Medicine
Essential Components Of Valid
Spirometry
Equipment Performance
Testing Technique
Measurement of Results
Technician Training
Interpretation Of Results
Selection of Reference Values
Race Adjustment of Predicted Values
Cross-sectional Evaluation: Normal, Obstructed,
Restricted
Changes Over Time
Pre- to Post- Bronchodilator Changes in Pulmonary
Function
Acute Work Related Changes in Pulmonary
Function
Summary
Appendix A - Glossary Of Terms And
Abbreviations
References
Lead Author & Committee
Spirometry in the Occupational Setting
This position paper provides the occupational physician with guidelines for
using spirometry testing in workplace medical programs. The focus is primarily
on conducting and interpreting spirometry tests in individual workers, though
spirometry data are also analyzed for groups of workers in respiratory
surveillance programs and epidemiologic research studies. The topics reviewed by
the statement are presented in Table 1. A glossary of
pulmonary function terms and abbreviations is provided in Appendix A.
Principles of Spirometry
Spirometry is the most basic and frequently performed test of pulmonary
function, measuring the ventilatory function of the respiratory system, i.e.,
the ability to move air into and out of the lungs. Using a forced expiratory
maneuver, which is a maximal expiration from Total Lung Capacity to Residual
Volume, spirometry measures volumes and flow rates. The expired air is measured
by a spirometer and the graphic recording of the expiration is called a
spirogram. For the past 50 years, volume-time spirograms have displayed expired
volume as a function of expiratory time (Figure 1). Since the
mid-1970's, flow-volume spirograms have also become common, showing expiratory
flow rate as a function of expired volume (Figure 2). As described below,
both displays are critical in assessing the technical quality of a test. Because
spirometry is based on a maximal, forced expiratory maneuver, the accuracy of
its results are effort-dependent, requiring a subject's full understanding,
cooperation, and effort.
Three clinically useful measurements are obtained from a properly performed
spirometry test. The Forced Vital Capacity (FVC) measures the total volume of
air exhaled during the maneuver. Speed of the expiratory airflow is quantified
by the Forced Expiratory Volume in One Second (FEV1 ), and by the
relationship of the FEV1 to the FVC, expressed as the FEV1 /FVC ratio. These
measurements are usually compared with average values "predicted" for a subject
based on their sex, age, height, and race. An FEV1 /FVC that is below the lower
limit of a subject's normal range for this ratio indicates probable airways
obstruction. The severity of obstructive impairment is determined by the degree
of FEV1 reduction relative to its normal range. In the absence of airways
obstruction, an FVC that is below the lower limit of a subject's normal range
suggests restriction of lung volume; the severity of restrictive impairment is
reflected by the degree of FVC reduction. In addition, changes in FVC and FEV1
can be measured over time, to determine whether loss of function is excessive.
However, the criteria for evaluating longitudinal changes in individuals are
less standardized.
An additional measurement, the Forced Expiratory Volume in Six Seconds (FEV6
), is currently under consideration as a surrogate for the FVC, particularly in
the screening setting. However, at the present time, few sets of predicted
values include the FEV6 , limiting its usefulness. As predicted values are
published for the FEV6 , it may become an easily standardized substitute for the
FVC in assessing impaired pulmonary function. It is important to note that the
FEV6 must be compared with a predicted FEV6, and not a predicted FVC.
Indications for Spirometry in Occupational Medicine
When used appropriately, spirometry can play an important role in the
primary, secondary, and tertiary prevention of respiratory disease in the
workplace (1).
In the primary prevention of respiratory disease, spirometry can be
used in pre-placement and fitness-for-duty examinations of individuals, in
situations where: a) the physical demands of a job require a certain level of
cardiopulmonary fitness, e.g., heavy manual labor or fire fighting; or b) the
characteristics of respirator use can impose a significant burden on the
cardiopulmonary systems, e.g., use of a self-contained breathing apparatus, or
prolonged use of certain negative-pressure masks under conditions of heavy
physical exertion and/or heat stress (2,3). Though not required routinely under
the OSHA Respiratory Protection Standard, 29 CFR 1910.134, spirometry may be
used in the evaluation of respirator users in some situations (2,3,4).
In addition to the pre-placement screening of individuals, primary prevention
of occupational respiratory disease also includes research and monitoring of
health status in groups of workers. Potential health effects are assessed in
occupational groups by comparing workers exposed to an agent or process with
those not exposed and/or those with varying levels of exposure. This aspect of
primary prevention is particularly important in occupational medicine to detect
previously unrecognized health consequences following occupational exposures to
specific agents.
In the secondary prevention of respiratory disease, repeated
spirometric evaluations can be used in medical surveillance programs when
workplace exposures put workers at risk of developing occupationally related
respiratory disorders (1). Surveillance is needed to detect the slowly
developing or delayed losses of function that characterize many work-related
respiratory disorders. In this case, many healthy individuals are tested to
detect early excessive declines in the pulmonary function of a subgroup of
sensitive workers, even though the spirometry test results of these workers may
still remain in the normal range.
Respiratory surveillance programs require that a baseline be established and
that workers be re-tested periodically. These periodic spirometry tests may be
mandated by OSHA regulations, e.g. for employees exposed to asbestos, cadmium,
coke oven emissions, or cotton dust, and for respirator-wearers exposed to
benzene, formaldehyde, or methylene chloride, or recommended by OSHA Special
Emphasis Programs, e.g. Silicosis. The contents of the OSHA-mandated physical
examinations are summarized in a 1998 publication from the U.S. Department of
Defense "Occupational Medical Surveillance Manual" (5). The National Institute
for Occupational Safety and Health (NIOSH) also recommends respiratory
surveillance for more than two dozen additional exposures that do not have
OSHA-mandated surveillance programs (6). Periodic spirometry tests may also be
part of industry- or company-mandated medical surveillance programs or a
component of workplace health promotion programs. As will be discussed later in
the statement, the limitations of spirometry must be borne in mind when
interpreting periodic spirometry test results in individuals. While spirometry
can detect large changes over a short time or smaller changes cumulated over a
longer observation period, spirometry is not sensitive to small short term
changes in an individual's pulmonary function.
In the tertiary prevention of respiratory disease, spirometry is used
in the clinical evaluation of symptomatic individuals, since many pulmonary
diseases manifest themselves as restrictive, obstructive, or combined
ventilatory defects. Spirometry allows some quantification of the severity of
lung function loss and is one of the pulmonary function tests used in assessing
respiratory impairment. Spirometry may be a required component in the evaluation
of workers for disability under the Social Security Administration (SSA) (7),
the Federal Coal Mine Health and Safety Act (CMHSA) (8), and in the workers'
compensation setting (9,10). Though mild spirometric abnormalities are "usually
not correlated with diminished ability to perform most jobs," "progressively
lower levels of lung function [are] correlated with diminishing ability to meet
the physical demands of many jobs (10)." Additional measures of functional
impairment, such as the determination of diffusing capacity for carbon monoxide
(11), measurement of lung volumes (12), exercise tolerance testing (13), or
methacholine challenge testing (14) are beyond the scope of this statement.
Essential Components of Valid Spirometry
Spirometry is simple but fraught with technical pitfalls that can invalidate
the pulmonary function measurements. Failure to obtain full understanding,
cooperation and effort from a subject during any part of the test usually
results in an underestimation of the true pulmonary function. Poorly
maintained spirometers also affect the accuracy of observed spirometric values
(15-17). Such erroneous measurements may cause a normal, healthy subject to be
mislabeled as "impaired" or lead to incorrect assessments of impaired subjects.
When evaluating changes over time, small decrements in pulmonary function may be
lost in the noise of the measurements if testing equipment and/or technique are
not as accurate, precise, rigorous, and standardized as possible (18). For
analysis of group data, small differences between groups, which may be
scientifically important, can be obscured by poor quality data caused by
inadequate testing technique.
In occupational medicine, the consequences of such misinterpretations can go
beyond simply making an inaccurate diagnosis; decisions regarding fitness for
duty, workplace accommodation, and compensation for work-related illness may
also be affected. Furthermore, since occupational spirometry tests are often
conducted in the regulatory and medical-legal arenas, the validity of the
spirometry test is likely to be scrutinized. Therefore, it is critical
for both clinical and administrative purposes that occupational medicine
physicians understand the need for standardization and quality control in
spirometry.
Although timed forced expirations have been measured since the 1950s (19), it
has only been in the past two decades that spirometry standardization and
quality control have been emphasized. The American Thoracic Society (ATS) has
been at the forefront of these efforts, with spirometry standardization
statements and updates issued in 1979 (20), 1987 (21), and 1995 (22), as well as
interpretation guidelines issued in 1991 (23). Recommendations for infection
control and hygiene during spirometry testing are included in the most recent
Spirometry Update (22), and current research supports the continued validity of
these recommendations (24-27).
As listed in Table 1 , validity
of spirometry tests is affected by four elements: 1) equipment performance; 2)
testing technique; 3) measurement of results; and 4) technician training. Though
the details of each of these topics are extensively discussed in the 1994 ATS
Spirometry Update (22) and in applicable regulations (7,8,28), some key aspects,
often not appreciated by the occupational health community, are highlighted
below.
Equipment Performance
As summarized in Table 2, spirometers
can be classified into one of two types, depending on their mechanical
characteristics: volumetric spirometers accumulate and directly measure
exhaled air volume as a function of time, while flow-type spirometers
indirectly measure airflow during exhalation and integrate the flows to obtain
expired volume (19,29-33). While volume and flow-type spirometers are
distinguished by their mechanical characteristics, it should be noted that both
types of spirometers can produce both volume-time and flow-volume spirograms if
the spirometry software is programmed appropriately.
In general, volumetric and flow-type spirometers each have advantages and
disadvantages. In a volumetric spirometer, the subject's expired air may: a)
cause a collection bell to rise in a water jacket (water-sealed spirometer); b)
displace a piston horizontally in a cylinder, causing the seal between the
piston and cylinder to roll on itself (dry rolling seal spirometer); or c) fill
a bellows (bellows spirometer). The air-collecting part of the spirometer often
has a direct pen linkage, inscribing a volume-time spirogram on moving chart
paper during a subject test. In general, volumetric spirometers are precise,
operate simply, and are easily maintained. The chief disadvantage of volume
spirometers is their size, since they must be able to accumulate 8 liters of
expired air.
Flow-type spirometers, on the other hand, are lightweight and portable, since
their components are small, but their mechanical operating characteristics are
complex because the measurement of expired volumes is indirect and the range of
flows to be measured during a forced expiration is large (33). Different
flow-type spirometers measure: a) pressure differentials created as expired air
passes through an orifice or across a resistance element, e.g., composed of
parallel capillary tubes or a mesh screen (pneumotachometer); b) rotation speeds
of a turbine as expired air flows across it (turbine); or c) electrical current
required to maintain the temperature of a heated wire as expired air flows
across it (hot wire anemometer). The relationship between the measured index,
i.e. pressure, turbine speed, or electrical current, and flow rate is not always
linear, and many flow sensors perform better at high flow rates, encountered
early in the forced expiration, than they do at low flow rates, seen at the end
of the maneuver, particularly in subjects with airways obstruction. Flow-type
spirometers, in general, exhibit more variability (less precision) than
volumetric spirometers, which can adversely affect interpretation of the serial
spirometry measurements of medical surveillance programs (33).
Since a flow-type spirometer sensor is designed to detect pressure, turbine
speed, or electrical current, and the transducer is calibrated to relate the
measured index to rates of airflow, the integrity of the sensor must be
maintained to achieve accurate measurements of pulmonary function. The
characteristics of the sensor may become modified during spirometry tests if the
sensor is damaged, blocked, or if moisture condenses on or mucus obstructs a
resistance element, turbine, or hot-wire. Such altered sensor characteristics or
other electronic problems may produce test results that are erroneous, e.g. flow
rates that exceed the maximum flow capability of the instrument, exhaled volumes
that far exceed those expected for the subject, or results that continually
improve during a test session for every subject tested. It is critical
that users be alert for such subtle indications of malfunction.
Unlike respirators, spirometers are not certified or approved by a government
or private agency. However, as shown in Table 3, for both
types of spirometers, the ATS recommends minimal performance criteria (including
size of graphical display), validation of spirometers to determine whether
specific models meet the performance criteria, and frequent quality control
(calibration) checks to insure that spirometers remain accurate during use (22).
The requirements of the SSA (7) and the CMHSA (8) differ from the ATS
recommendations in some details; these regulations should be consulted prior to
conducting spirometry tests for impairment/disability evaluations.
The 1994 ATS Update presents a spirometer testing protocol for validating the
accuracy and precision of each spirometer model (22). This testing can be
performed by a spirometer manufacturer or by an independent testing laboratory.
The validation protocol uses standard waveforms (34) to drive a mechanical
syringe, delivering known volumes at known speeds into the spirometer and
software to be tested (35). The 1994 ATS Update testing protocol is far more
rigorous than previous ATS recommendations, so users should be certain that
their spirometer was tested using the 1994 ATS protocol. American College
of Occupational and Environmental Medicine (ACOEM) recommends that users request
written verification from the manufacturer indicating that a particular
spirometer has successfully passed its validation checks, and that the tested
spirometer and software version correspond with the model and software version
that is being purchased. However, it must be stressed that such validation under
laboratory conditions does not guarantee that a device will retain its accuracy
and precision under field conditions; the importance of frequent calibration
checks in the field cannot be overstated.
Even when spirometers meet the minimal criteria set out by the ATS, they
still vary in the accuracy and precision with which they measure expired volumes
of air, in the completeness of the visual display presented to the technician
for recognizing testing errors, in the availability of extensive
computer-derived technical quality indicators (36-39), in the information that
is saved as a testing session progresses and after the session is completed, and
finally, in whether data points from tracings are saved so that the tracings can
be recalled at a later date for comparison with other tests or for quality
control reviews of spirograms (Table 3). The best
systems far exceed ATS recommendations for accuracy and precision, provide
real-time visual displays of the expiratory maneuver as well as computer-derived
technical quality indicators, store all information from a test session, and
save data points so that tracings can be reconstructed electronically at a
future time. Users must remember that the highest degree of precision and
accuracy is needed when serial spirometry measurements will be evaluated for
small changes over time.
Unless the spirometry system saves electronic copies that permit whole
spirograms from past test sessions to be displayed or printed, ACOEM recommends
that hard copies of tracings should be maintained so that the technical quality
of tests can be examined when necessary. This is particularly important in the
case of clinics and practices providing occupational health services, where
providers of medical services may change periodically. The capability of
examining volume-time curves to check the end of test and flow-volume curves to
check the beginning of exhalation is essential in determining whether spirometry
test results are probably valid or reflect obvious testing artifacts (22).
Calibration tracings and records support the validity of spirometry tests
conducted on a particular day with a particular spirometer. Since OSHA requires
that medical records be retained for 30 years after termination of employment
(40), ACOEM recommends that these calibration records should be saved and a log
kept of any problems found and solved or any changes in protocol, computer
software, or equipment that were made. Thermal paper should be photocopied since
it fades rapidly over time.
It is important to note that a new National Institutes of Health-sponsored
program, the National Lung Health Education Program (NLHEP), is being developed
to encourage primary care physicians to screen smokers for Chronic Obstructive
Pulmonary Disease (COPD) (41). NLHEP requires less rigorous testing procedures
and documentation than are required for occupational spirometry testing, as well
as encouraging the use of new inexpensive "office spirometers." Occupational
medicine physicians need to be cautioned that many of NHLEP's testing procedures
and "office spirometers" are not acceptable for diagnostic spirometry or for
occupational screening, surveillance, and impairment evaluations.
Testing Technique
OSHA (28), SSA (7), CMHSA (8), and the ATS (22) make specific recommendations
regarding performance of the forced expiratory maneuver and measurement of the
spirogram. Key elements from the 1994 ATS Update and changes from the 1987 ATS
guidelines are summarized below and in Table 4.
Testing should be conducted at ambient temperatures between 17 - 40° C.
However, temperatures > 23° C are preferable to avoid a large
temperature difference between the spirometer temperature and body temperature
(42). If a large difference exists, the exhaled air cannot fully cool down to
the spirometer temperature within the first second of exhalation. In this case,
an inappropriate correction factor, based on the spirometer temperature, will
usually be selected to adjust the exhaled volume from spirometer to body
temperature (BTPS correction), causing inflated measurements of BTPS-corrected
FEV1 (42,43).
The technician must demonstrate correct performance of a spirometry
test, as well as describing it verbally, to the subject being tested. The
technician must enthusiastically coach the subject to record "acceptable"
maneuvers, which have good starts, are free from artifacts, and have
satisfactory exhalations (Table 4).
Specifically, the subject must: a) exhale with a hard fast "blast" of air so
that the volume of air leaked out before the blast (the "extrapolated volume")
is less than 5% of the FVC or 0.150 L, whichever value is greater; b) exhale
smoothly, with no cough or glottis closure in the first second, and no leak,
obstruction of the mouthpiece, or variable effort; and c) exhale completely, for
at least 6 to 10 seconds and/or until a one second long FVC plateau is reached,
unless the subject cannot exhale this long because of discomfort, airways
obstruction, or advanced age.
The testing goal is to record at least 3 acceptable maneuvers with the
best FVC and the best FEV1 reproduced to within 0.20 L, attempting up to 8
maneuvers if necessary (22). Failure to meet these criteria does not rule out
interpretation of results, since some impaired subjects may have difficulty in
attaining them (44-46). However, when interpreting such results, it must be
borne in mind that tests failing to meet the testing goal usually
underestimate true pulmonary function.
The need for electronic or hard copies of a test session to support the
"acceptability" of the test session cannot be overstated. Adequacy of the end of
test is best checked by examining volume-time curves for evidence of an FVC
plateau and length of expiration (Figure 1 ); the
beginning of exhalation is best checked by examining flow-volume curves from
each maneuver for an immediate rise to a sharp peak in expiratory flow rate
(Figure 2).
Unacceptable spirograms are depicted in the 1994 ATS Spirometry Update (22) and
in some reference books (30). Examination of hard copy or electronic tracings is
probably the only way of evaluating whether trends in spirometry test results
may be real or obviously reflect testing artifacts. ACOEM strongly recommends
that hard copies and/or electronic copies of spirograms be saved from spirometry
test sessions.
Measurement of Results
The largest FVC and the largest FEV1 from the acceptable curves are reported
for a subject, even if they are not derived from the same maneuver
(
Table 5). Also, the
largest FEV1 may come from a curve that is acceptable except for its early
termination (22). All expiratory flow rates are drawn from the single acceptable
tracing having the largest sum of FEV1 FVC. Users should check their spirometers
to ensure that their spirometry software selects the correct values for the test
report. All observed volumes and flow rates are corrected to body temperature
(BTPS).
Technician Training
In 1978, OSHA prescribed elements of standardization for spirometry in the
occupational setting when it promulgated the Cotton Dust Standard, 29 CFR
1910.1043 (28). The need for technician training is emphasized in the Preamble
to the Standard: "The key to reliable pulmonary function testing is the
technician's way of guiding the employee through a series of respiratory
maneuvers. The most important quality of a pulmonary function technician is the
motivation to do the very best test on every employee. The technician must also
be able to judge the degree of effort and cooperation of the subject. The test
results obtained by a technician who lacks these skills are not only useless,
but also convey false information which could be harmful to the employee."
Based on the "Qualifications of personnel administering the test" given in
Appendix D of the Cotton Dust Standard, NIOSH developed a program that reviews
and approves spirometry training courses. Cotton Dust Standard Appendix D
outlines the content of NIOSH - approved spirometry courses and states that the
goal of these courses is to provide technicians with "the basic knowledge
required to produce meaningful results." For many exposures, OSHA requires that
technicians attend courses "sponsored by an appropriate academic or professional
institution" or a NIOSH-approved course (28,47,48). Though attendance at a
NIOSH-approved course is not required for technicians outside of the cotton
industry, most companies view NIOSH approval as minimal assurance that the
course will adequately teach the basic principles of spirometry. NIOSH currently
approves about one course per year; 50 courses that have been approved are
currently active. ACOEM (49), NIOSH (43), ATS (22), and the American Association
of Occupational Health Nurses (AAOHN) (50) all recommend technician training to
ensure accurate pulmonary function testing.
Spirometry refresher classes are not mandated by any OSHA regulations, nor
does NIOSH approve the content of refresher courses. However, the need for
repeated training of technicians was recognized and documented in the National
Institutes of Health-sponsored multi-center Lung Health Study (37) and the
NIOSH-monitored spirometry of the National Health and Nutrition Examination
Survey III (NHANES III) (38), and ACOEM has recommended "periodic, e.g. every 3
years," refresher courses for many years (49). Spirometry refresher courses keep
technicians informed of changes in occupational pulmonary function testing, and
reinforce the need for vigilance in conducting spirometry tests. Technician
drift and apathy develop if no feedback is provided on test quality, and on the
importance of active coaching and recognition of testing errors. Intensive
refresher courses designed for experienced technicians are recommended, rather
than attending part of a NIOSH-approved spirometry course.
The 1994 ATS Spirometry Update strongly emphasizes the importance of
technical quality in achieving valid spirometry results; figures of many
technical errors that plague spirometry testing are presented in the Update
(22). ATS recommends that spirograms be reviewed periodically, to provide
regular feedback on the quality of each technician's testing. Quality control
reviews can be performed on tracings that are saved electronically during the
testing session, or on photocopies of randomly selected spirograms.
As summarized in
Table 5, ACOEM
strongly recommends that spirometry technicians in the occupational setting
complete a NIOSH-approved spirometry course as part of their training.
Increasingly, clinics and practices engaged in providing occupational medical
services may argue that such training is not needed for adequate performance of
the test. However, recognition of the technical pitfalls of spirometry is
critical in the occupational area, and NIOSH-approved courses are specifically
geared toward training technicians to conduct screening spirometry tests,
recognizing these pitfalls. In addition, ACOEM continues to recommend that
technicians attend spirometry refresher courses every 3 years, to discuss
testing problems. Such courses encourage technicians to remain vigilant and
enthusiastic during spirometry testing of workers. If feasible, a program
providing quality assurance review of spirograms is also highly recommended.
Interpretation of Results
Interpretation of spirometry results should always begin with an
assessment of test quality (22). Once the validity of the measurements has been
established, the evaluation of the test subject's lung function can proceed.
Interpretation of results is summarized in Tables 6 and
7 and
Figure 3
Selection of Reference Values
The first step in interpreting pulmonary function results is usually to
determine where the subject's spirometry values fall relative to the normal
range. Ideally, this normal range would be based on a population similar to the
workers being examined, with spirometry measurements made and analyzed in
accordance with the most recent ATS recommendations, using equipment and testing
technique similar to that employed in testing the workers under consideration
(23). However, reference "predicted" values that define the normal range are
often drawn from relatively small numbers of subjects resident in a single
geographic location, often near or accessible to an interested research
investigator. Reference values may be derived from an institutional or
occupational group, a population-based epidemiologic study, or subjects chosen
specifically to create reference equations (23). Within the study group, the
relationship between pulmonary function and age, height, and sex is summarized
in regression equations, which are usually named after the primary investigator.
In clinical medicine, many laboratories use the equations of Morris (51), Crapo
(52), or Knudson (53), depending to some degree on which equations are
programmed into the automated spirometry equipment (54).
In the occupational setting, Knudson's prediction equations have been widely
used because the 1976 equations (55) were mandated by the OSHA Cotton Dust
Standard: they were the only equations available at the time that studied both
males and females, were based on non-smokers, and used back-extrapolation to
define time zero. The Knudson data were re-analyzed in 1983 (53) using data
selection criteria that conform to ATS recommendations, resulting in equations
that predict considerably different values than in 1976, particularly for the
forced expiratory flow rates. Crapo's prediction equations (52) are also used in
the occupational setting since they were adopted by the American Medical
Association (AMA) as the standard reference in the 4th Edition of the
AMA Guides to the Evaluation of Permanent Impairment (9). Many reference
equations are listed in the ATS Interpretative statement (23), and the 1997
NIOSH Spirometry Training Guide demonstrates the varying results obtained when
different prediction equations are used (43). As recommended by the ATS, the fit
of a set of reference values to a particular occupational setting can be checked
empirically by testing 20-40 local non-smoking healthy subjects and determining
their percentages of predicted using the intended reference equations (23).
It should be noted that an important alternative source of spirometry
reference values has recently become available for both the clinical and the
occupational settings. In January, 1999, race/ethnic group-specific equations
were published from the third National Health and Nutrition Examination Survey
(NHANES III), based on a random sample of the U.S. population using standardized
state-of-the-art spirometry testing methodology (56). The NHANES III data
permitted reference equations to be calculated separately for Caucasians,
African-Americans, and Hispanics. Though a few regulations and guidelines
continue to require the use of specific sets of reference values (8,9,28), ACOEM
recommends that the NHANES III equations be considered for general use in the
occupational setting as these equations become available in computerized
spirometry systems.
Race Adjustment of Predicted Values
Publication of the NHANES III prediction equations is an important step
forward, not only because the reference values are based on a random sample of
the U.S. population that was examined in the last few years, but also because
predicted values specific for African-Americans and Hispanics, based on randomly
selected subjects from the U.S. population, are now available. Until this time,
the most widely used reference values have been derived from Caucasian
populations in North America. Prior to 1978, when workers in the cotton industry
were evaluated using these Caucasian reference values, more abnormal spirometry
results were noted among African-American than among Caucasian workers. Since
race-specific reference equations were not in general use in 1978, OSHA mandated
that "the predicted FEV1 and FVC for blacks should be multiplied by 0.85 to
adjust for ethnic differences" (Table 6).
At the time, OSHA recognized that "this correction may not be precisely
correct," but OSHA relied on the current state of the art "to provide proper
interpretation of spirometry measurements for blacks without inadvertently
fostering discrimination in hiring practices (28)." The practice of adjusting
Caucasian predicted values for FVC and FEV1 for African-American subjects has
remained widespread in the occupational setting since 1978. However,
race-adjustment is less widely used in the clinical setting (54).
The 1991 ATS Official statement on "Lung Function Testing: Selection of
Reference Values and Interpretative Strategies" recommends use of race-specific
prediction equations such as the NHANES III (56) if "possible and practical," or
cautious use of a scaling ("race-adjustment") factor of 0.88 if non-Caucasians
are tested infrequently ( 23). It is important to use a subject's
self-declared race or ethnic group as a basis for selecting appropriate
race-specific predicted values or for deciding whether or not to race-adjust
Caucasian predicted values. Though using race-adjusted Caucasian predicted
values for African-American subjects is preferable to using non-adjusted
Caucasian predicted values (57), recent studies conclude that a single
adjustment factor is not optimal and that race-specific equations should be used
(57,58).
There is less consensus on the adjustment of Caucasian predicted values for
other ethnic groups, such as Hispanics, Asians, and Pacific Islanders than there
is for African-Americans (59). Current sources and studies do not recommend
race-adjustment for any of these groups except some Asians, i.e. Chinese and
Japanese, in addition to African-Americans (37,59).
As noted above, ACOEM recommends that occupational settings consider adopting
the NHANES III equations for general use as they become available in spirometry
systems. Until these equations are available, ACOEM recommends that Caucasian
predicted values should be race-adjusted for African-American, Chinese, and
Japanese subjects, applying the ATS recommended scaling factor of 0.88 to the
Caucasian predicted FEV1 and FVC (Table 6).
However, if testing is conducted under the few regulations and guidelines that
have specific recommendations/requirements regarding race-adjustment factors
(8,9,28), those requirements should be followed.
Cross-sectional Evaluation: Normal, Obstructed, Restricted In its 1991
Interpretation statement, the ATS recommends that spirometry results be
interpreted based on a stepwise algorithm using very few parameters (23,30), as
summarized in Figure 3.
A value of the FEV1 /FVC % of predicted below the Lower Limit of Normal (LLN)
indicates probable obstructive impairment. Having established the presence of
obstruction, the FEV1 % of predicted is used to grade the degree of obstructive
impairment. There are several definitions of severity categories available
(9,10,23,30), and Figure 3 presents the ATS respiratory impairment categories
(10), which define "mild" obstruction as an FEV1 between 60% of predicted and
the LLN, "moderate" obstruction as an FEV1 of 41-59% of predicted, and "severe"
obstruction as an FEV1 of 40% or less of predicted. "Borderline" obstruction
may exist when a subject's FEV1 /FVC % of predicted is below its LLN, but
the FEV1 falls within the normal range. However, the ATS cautions that "the
pattern of a low FEV1 /VC ratio and greater than average VC and FEV1 should be
recognized as one that may occur in healthy individuals" (23).
In the absence of airways obstruction, the FVC % of predicted is used to
determine whether restrictive impairment is present, with the ATS defining
"mild" restriction as an FVC between 60% of predicted and the LLN, "moderate"
restriction as an FVC of 51-59% of predicted, and "severe" restriction as an FVC
of 50% or less of predicted (10).
Contrary to long-standing practice, the use of a fixed cut-off of 80% of
predicted as a LLN is not recommended (Table 7),
and should be replaced by the fifth percentile, the point below which 5% of
normal subjects fall (23). The LLN should be obtained from the same source as
the predicted values, from tables or equations presented in the reference (53,
56) or calculated as: LLN = 1.645 x SEE (23). LLNs calculated in this way tend
to decline with age, and thus can have an impact on whether a 50-60 year old
employee is labeled as "normal" or "abnormal." For example, using the 1983
Knudson prediction equations (53), the LLN (5th percentile) for FVC for a man of
40 years or older is 73.4% of predicted, which is significantly below the
previously used 80% of predicted. Finally, due to wide variability within and
between healthy subjects, the ATS states that "FEF25-75% and the instantaneous
flows should not be used to diagnose small airway disease in individual
patients" (23) or to assess respiratory impairment (10). Interpretation of FEF
25-75% and other flow rates is not recommended if the FEV1 and the FEV1 /FVC are
within the normal range, though the flow rates "may be used to confirm the
presence of airway obstruction in the presence of a borderline FEV1 /VC" (23).
In other words, an FEF25-75% % of predicted below its LLN can confirm the
presence of airways obstruction in subjects falling into the "Possible
Borderline Airways Obstruction" category in Figure 3.
However, such interpretations should bear in mind the ATS's warning that a low
FEV1 /FVC ratio accompanied by FVC and FEV1 that are above average, i.e. >
100% of predicted, can occur in healthy individuals (23).
The degree of variability in the FEF25-75% is reflected in the low value of
its 5th percentile LLN. Using the 1983 Knudson prediction equations, the 5th
percentile LLN for FEF25-75% for a man of 40 years or older is 40.3% of
predicted, indicating that a man over 40 must be less than half of his Knudson
predicted value before he falls below the normal range.
Changes Over Time
In the occupational setting, changes over time in pulmonary function should
be examined for two reasons: a) to evaluate a worker's response to treatment in
the clinical setting, and b) to screen healthy workers for excessive loss of
function over time. In the first situation, the ATS recommends a non-algorithmic
approach to interpretation, stating that "the clinician seeing the patient can
often interpret results of serial tests in a useful manner, not reproducible by
any simple algorithm. For example, seemingly stable tests may prove very
reassuring in a patient receiving therapy for a disease that is otherwise
rapidly progressive. The same tests may be very disappointing if one is treating
a disorder that is expected to improve dramatically with the therapy prescribed.
Depending on the clinical situation, statistically insignificant trends in
function may be very meaningful to the clinician (23)."
The second situation, screening healthy workers for excessive loss of
pulmonary function, is often encountered in workplace medical surveillance
programs. When subjects' spirometry test results are compared with a
cross-sectional LLN, as described in the previous section and shown in Figure 3,
excessive loss of pulmonary function will be identified adequately in workers
with average or less than average lung size. However, such evaluations will not
detect early excessive loss of function in workers whose lung size is above
average, i.e., above 100% of predicted. Particularly for these subjects, change
in pulmonary function over time should be included in a screening program to
determine whether the worker's spirometry test results are decreasing faster
than expected over time (23,36,60).
Loss of FEV1 or FVC over time can be estimated simply by calculating the
difference between volumes measured at two points in time, or by fitting a least
squares "slope" through periodic measurements over time for an individual. Since
estimates of individual rate of change become more precise as follow-up time
increases, loss of FEV1 or FVC should be estimated from measurements made over a
minimum of 4-6 years (61-64). Measurement frequency has less impact on
precision than length of follow-up does (61,62), but periodic measurements are
needed to detect workers experiencing rapid declines in pulmonary function and
to detect systematic differences between examinations over time (62,64).
Interpretation of change over time in the screening setting is complicated by
the substantial variation in rates of change that exists both between workers
and within an individual worker. Though the FEV1 and FVC can be measured
precisely during one test session, biological and technical variation over time
make an individual's estimated rate of change over time highly variable (61-65).
Though within-day variability for a normal subject's FEV1 and FVC is <
5%, year-to-year variability is < 15% (23,62). Technical variability
can be minimized by using very precise spirometers, not changing equipment
unnecessarily over time, and maintaining a rigorous spirometry quality assurance
program; biological variability can be reduced by conducting successive
spirometries at about the same time of day and in the same month each year.
Because of the precision gained by combining results from many subjects, group
estimates of change can be calculated and comparisons made between groups in
epidemiologic studies.
Epidemiologic data have indicated that for adult smokers "to develop
clinically significant airflow obstruction, the average rate of decline of FEV1
… probably needs to be greater than 90 ml/year, or about three times that seen
in non-smokers and twice the rate seen in the 'non-susceptible' smokers" (65).
One study found that about 4% of their combined smoking and non-smoking male
population had FEV1 slopes of 100 ml/year or greater when measured over 4-11
years of follow-up (66). However, studies differ in their estimates of change
over time, and, to date, neither longitudinal predicted values nor 5th
percentile LLNs have been recommended for the evaluation of individual rates of
change over time in occupational or clinical settings (62).
To meet the need for longitudinal LLNs, the ATS recommends a conservative
strategy to minimize false positives in the screening setting, stating that:
"The greatest errors occur when one attempts to interpret serial changes in
subjects without disease because test variability will usually far exceed the
true annual decline, and reliable rates of change for an individual subject
cannot be calculated without prolonged follow-up. Thus, in subjects with
'normal' lung function, changes in VC or FEV1 over 1 year should probably exceed
15% before any confidence can be given to the opinion that a meaningful
year-to-year change has occurred (23)." NIOSH adopted this definition of
significant change in a 1995 Criteria Document, stating that "because of
considerable short-term variability in FEV1, a year-to-year change of greater
than 15% should occur before a change in FEV1 is considered significant." NIOSH
concluded that "evidence of impaired lung function is present when there is a
confirmed finding of a decline in FEV1 (adjusted for the expected interval
decline in FEV1 ) of 15% or greater" and that such a decline "is considered
significant and warrants further medical evaluation" (70).
Since FEV1 and FVC decline with age from the about the mid-30's on, with some
acceleration of the rate as aging advances (67,68), an allowance for the
expected loss due to aging should be made before labeling a 15% decline as
"significant"(36,70). As Appendix G of the NIOSH Criteria Document (70) states:
"The LLN for the follow-up FEV1 is computed by taking 85% of the baseline value
minus the expected decline over the time period. An individual's expected
decline over the time period is dependent on his/her age, but for practical
considerations, a constant value of 25 ml/year is often recommended. For
example, an individual whose initial FEV1 is 4.00 L would be considered to have
an accelerated decline in FEV1 if his/her FEV1 is below 3.15 L, 10 years after
the baseline value was determined [(0.85 x 4.0 L) - (10 years x 0.025 L/year) =
3.15 L]." Such a loss over 10 years would be labeled "significant," and would
warrant medical evaluation once the low value was confirmed by a re-test
(36,70).
In summary, as shown in Table 8,
ACOEM recommends that spirometry should be conducted every 1-2 years when
indicated because of workplace exposures, unless otherwise specified by
applicable regulations or recommendations. The frequency of testing may vary
with age and length of exposure as in the National Fire Protection Association
(NFPA) examination protocol, which recommends spirometry testing every 3 years
for fire fighters under age 29, every 2 years for ages 30-39, and annually for
ages 40 and above (69). Change in FEV1 and FVC over time should be evaluated as
part of a screening program once measurements have been made over at least 4-6
years. A decrease in FEV1 or FVC of 90-100 ml/ year, over at least 4-6 years,
should trigger further scrutiny of a worker's pulmonary function measurements
over time. Loss of 15% or more of the observed FEV1 or FVC, after allowing for
the expected decrease due to aging, should be regarded as a significant decline
over time. If the low results are confirmed on a re-test, a medical review is
warranted, even if the worker's values still remain above the cross-sectional
LLN.
Pre- to Post- Bronchodilator Changes in Pulmonary Function
The ATS (23,71) and the NHLBI National Asthma Education and Prevention
Program (NAEPP) (72) recommend that a pre- to post-bronchodilator increase in
FEV1 should be at least 12% of initial FEV1 and at least 0.2 liters to be called
significant, i.e. a bronchodilator response that is suggestive of airways
hyperreactivity (Table 8).
The Global Initiative on Asthma (GINA) (73) and the NHLBI Lung Health Study (74)
regarded a 15% increase in FEV1 as significant.
Attention should be limited to changes in the FEV1 because interpreting
changes in the FVC or FEF25-75% is likely to be complicated by varying lengths
of expiration recorded before or after the bronchodilator (23). If changes in
the FVC are examined, the ATS recommends that a change of at least 15% of
initial FVC be considered significant, i.e. suggestive of airway reactivity. The
ATS does not endorse interpretation of pre- to post-bronchodilator changes in
the FEF25-75% (23).
Based on these sources, ACOEM recommends that a pre- to post-bronchodilator
increase in FEV1 should be at least 12% of initial FEV1 and at least 0.2 liters
to be considered significant, i.e., suggestive of reversible obstructive airways
disease. However, it should be noted that failure to achieve such a response to
bronchodilators in the laboratory does not completely exclude the possibility of
reversible airways disease. ACOEM also concurs with ATS (10) and AMA (9) that
impairment determinations should utilize a worker's best values for FVC and FEV1
,whether recorded before or after bronchodilator administration.
Acute Work-Related Changes in Pulmonary Function
Work-related bronchoconstriction, causing decreased pulmonary function across
a work shift or increased variability in pulmonary function across a longer
period at work, can be elicited by bronchial irritants and sensitizers and is
often reversible. Patterns of work-related change are an important element in
the diagnosis of a number of occupationally related respiratory disorders,
particularly occupational asthma (Table 9
). Spirometry measurements should be made as close to the work environment as
possible to avoid a long time lapse between the worker's occupational exposure
and the measurement of pulmonary function. As discussed below, when occupational
asthma is suspected, additional measurements should also be made at home at the
conclusion of the work day to capture any delayed work-related declines in
function. The spirometry measurements most commonly examined are the FEV1 and
the Peak Expiratory Flow Rate (PEF or PEFR), though interpretation of FEV1
decline is better standardized than interpretation of PEF variability. Newly
marketed portable spirometers are becoming available for serial spirometry
measurements in the workplace, in addition to the traditionally used peak flow
meters.
In 1978, the OSHA Cotton Dust Standard defined an across-shift decrease in
FEV1 of 5% or 0.2 L, whichever is greater, as a significant drop if it is
confirmed within a month (28). In 1986, a drop in FEV1 of 5% or 0.15 L,
whichever is greater, was labeled as significant, if it is confirmed on a second
occasion (60). An FEV1 decrease of 10% would be considered significant if only
one pre- to post- shift study was performed (60,75).
When considering such small declines in FEV1 as significant, it is critical
to maintain the testing environment and the spirometer at 23° C (73° F) or above
(42,43). FEV1 declines of several percent can be produced as an artifact if the
testing environment and the spirometer warm up by several degrees between the
pre-shift and post-shift tests. Usually the BTPS correction factor is selected
based on the spirometer temperature, and not the temperature of the accumulated
exhaled gas one second after the expiration commences. With a cold pre-shift
spirometer, a large BTPS correction factor may be applied to exhaled air that is
still closer to body than to spirometer temperature, resulting in an inflated
"observed" FEV1 . With a warmer post-shift spirometer, the temperature of the
accumulated exhaled air is closer to the spirometer temperature so that the
selected BTPS factor is appropriate. Subsequently, calculation of a pre- to
post-shift change in FEV1 finds an FEV1 decline that depends on the warming of
the spirometer across the work shift rather than employee exposures in the work
environment (43).
Based on the sources described above, ACOEM recommends that a single pre- to
post-shift study finding a decline in FEV1 of at least 10% is significant and
worthy of follow-up. In addition, an across-shift decline in FEV1 of at least 5%
or 0.2 L, whichever is greater, that is confirmed on a second occasion, should
be reviewed and followed up. However, when only a few shifts are monitored,
declines in FEV1 of about 5% must be interpreted cautiously since this amount of
variation can be seen within a day in normal subjects (23).
Unlike the other spirometry evaluations discussed in this position paper,
serial PEF monitoring currently is used not as a screening test, but rather to
confirm suspected associations between a worker's respiratory symptoms and
exposures on the job, and to identify potential triggering exposures. Monitoring
PEF in relation to work exposures and respiratory symptoms is more complex than
examining pre- to post-shift changes in FEV1 because of the PEF's variability
and effort-dependence; a long time period and many measurements made during
periods of work exposure and periods away from work are needed to document
patterns of PEF (76). Although there is no uniformly accepted protocol,
comparable to the 1994 ATS Spirometry Update, for conducting and evaluating
serial PEF measurements in relation to workplace exposure, the protocols
recommended by the American College of Chest Physicians Consensus statement
(77), the NAEPP (72), the European Respiratory Society (78), and the
Subcommittee on Occupational Allergy of the European Academy of Allergology and
Clinical Immunology (79,80) generally agree that: a) every two hours, or at
least four times each day, three PEF measurements should be made and the highest
of the three analyzed, with one measurement made on first awaking and another
made close to the midpoint of the waking day to capture the worker's lowest and
highest PEFs (81-84); and b) if possible, workers should be monitored for at
least two weeks at work and two weekends to 10 days away from work, as needed,
to identify or exclude work-related changes in PEF (72,77-79). Interpretation of
PEF measurements can be confounded by use of bronchodilator or steroid
medications and by the normal diurnal variation in airway caliber, which
produces lower flow rates upon awaking and higher flow rates 6-8 hours later.
This normal pattern can mask the workplace-related decrement of an immediate
reaction in a day worker (85).
PEF measurements can be evaluated by calculating their daily variability,
with a mean diurnal variation of 20% or more probably indicating asthma
(72,77,79,81). A more sensitive, specific, and certainly more difficult
technique is to visually inspect graphs of the maximum, mean, and minimum daily
PEFs. These graphs, when examined by experienced interpreters, can be used to
investigate patterns of expiratory flow rates during workplace exposures and
during the time away from work (72,76-78).
Major factors can interfere with interpreting serial PEF measurements: a)
reliance on self-reporting by the monitored worker, requiring the worker's
motivation and honesty - though development of electronic peak flow meters that
store measurements, symptoms, and medication information may improve compliance
(72,76,78); b) occurrence of intermittent exposure to suspect agents in the
workplace; c) delay in making the first PEF measurement after waking on days
away from work; and d) use of more than one PEF meter (72,76). The 1994 ATS
Update provides specifications for the accuracy and precision of peak flow
monitoring devices (22), and peak flow meters have been evaluated (86). However,
visual interpretation of PEF patterns over time requires experience, and work is
ongoing to quantify the qualitative judgments that are made (76,78).
Summary
This position paper reviews several aspects of spirometric testing in the
workplace, where spirometry is employed in the primary, secondary, and tertiary
prevention of occupational lung disease. Primary prevention includes
pre-placement and fitness-for-duty examinations as well as research and
monitoring of health status in groups of exposed workers; secondary prevention
includes periodic medical screening of individual workers for early effects of
exposure to known occupational hazards; and tertiary prevention includes
clinical evaluation and impairment/disability assessment.
For all of these purposes, valid spirometry measurements are critical,
requiring: documented spirometer accuracy and precision, a rigorous and
standardized testing technique, standardized measurement of pulmonary function
values from the spirogram, adequate initial and refresher training of spirometry
technicians, and ideally, quality assessment of samples of spirograms.
Interpretation of spirometric results usually includes comparison with
predicted values and should also evaluate changes in lung function over time.
Response to inhaled bronchodilators and changes in relation to workplace
exposure may also be assessed. Each of these interpretations should begin with
an assessment of test quality and, based on the most recent ATS recommendations,
should rely on a few reproducible indices of pulmonary function (FEV1 , FVC, and
FEV1 /FVC.) The use of forced expiratory flow rates, e.g., the FEF25-75% , in
interpreting results for individuals is strongly discouraged except when
confirming borderline airways obstruction. And finally, use of serial PEF
measurements is emerging as a method for confirming associations between reduced
or variable pulmonary function and workplace exposures in the diagnosis of
occupational asthma.
Throughout the statement, ACOEM makes detailed recommendations to ensure that
each of these areas of test performance and interpretation follows current
recommendations/standards in the pulmonary and regulatory fields.
Acknowledgment: The authors would like to thank Drs. John L. Hankinson and
Robert O. Crapo for their many helpful suggestions and comments on this position
paper. In addition, we want to thank other members of the American Thoracic
Society, who shared their views and insights as this position paper was
developed.
Appendix A - Glossary Of Terms And Abbreviations
ATPS. Ambient temperature and pressure saturated with water vapor.
Volumes read directly off the volume-time spirogram are at ATPS.
Back extrapolation. In the calculation of FEV1, a method for determining
the time zero. A straight line is drawn through the steepest portion of the
volume-time curve back to the baseline. Where this straight line intersects the
baseline is the zero point for timing the FEV1 .
Best curve. That curve which gives the largest sum of FEV1 and FVC. The
best curve is used in the calculation of the FEF25-75% and the instantaneous
flow rates. In contrast, the largest FVC and the largest FEV1 are reported for
the test session, even if they are not from the same curve.
BTPS. Body temperature and pressure saturated with water vapor. All
spirometric volumes and flows must be corrected to BTPS.
Calibration check. Periodic determination of a spirometer's ability to
accurately measure volume. Calibration checks should be performed at least daily
using a three liter syringe. The instrument should maintain an accuracy of 3% of
the reading. Additional checks include checking for leaks (daily for volume
spirometers), and, every 3 months, verifying the accuracy of a timed chart and
checking the linearity of volume recording.
End of test. That point during the forced expiratory maneuver when a
plateau at least one second long is noted on the volume-time tracing
Extrapolated volume. That volume determined by a line drawn through the
zero time point perpendicular to the baseline on a volume-time curve. The
extrapolated volume is read where this perpendicular line intersects the volume
curve; it should be less than 5% of the FVC or 150 ml, whichever is
greater.
FEV1 /FVC%. Forced expiratory volume in one second expressed as a
percentage of the forced vital capacity.
Flow-measuring spirometer. Indirectly measures volume of exhaled air by
measuring the rate at which air is exhaled and deriving the volume. Examples
include pneumotachometer, mass flow, and turbine instruments.
Forced expiratory Volume in one second (FEV1 ). Volume of air exhaled
during the first second of the FVC.
Forced expiratory Volume in six second (FEV6 ). Volume of air exhaled
during the first six seconds of the FVC. Since it is easier for obstructed
subjects to reach the FEV6 than the FVC, there is growing interest in measuring
the FEV6 and the FEV1 /FEV6 in screening spirometry.
Forced expiratory maneuver. Technique during spirometry where the subject
takes the deepest possible inspiration from a normal breathing pattern and blows
into the mouthpiece as hard, fast and completely as possible. Also known as the
forced vital capacity maneuver.
Forced Vital capacity (FVC). The maximal volume of air exhaled from the
point of maximal inspiration using a maximally forced expiratory effort.
Mean forced expiratory flow during the middle half of the FVC (FEF25-75% ).
Average flow rate over the middle half of the expiration. Formerly called the
maximal mid-expiratory flow rate (MMEF).
Predicted normal values. Expected values for various lung volumes and
flow rates derived from healthy populations.
Reproducibility. In the absence of disease-related changes, the ability
of a test to obtain the same result from an individual repeatedly tested over a
period of time. Reproducibility of the FEV1 and FVC within a test session should
be 0.20 liters or less.
Residual Volume. Volume remaining in the lungs following a maximal
expiration.
Spirogram. A graphic recording of a forced expiratory maneuver, as either
a volume-time or flow-volume tracing.
Spirometer. An instrument for measuring lung volumes and flow
rates.
Total Lung Capacity. Total lung volume following a maximal
inspiration.
Valid Test. A spirometry test consisting of at least three acceptable
forced expiratory tracings where the best FVC and the best FEV1 are reproduced
within 0.2 L.
Volume-measuring spirometer. Spirometers which directly accumulate and
measure the volume of exhaled air as a function of time. Examples include
water-seal, dry rolling seal and bellows instruments.
Zero time point. In the measurement of FEV1 , the point selected as the
start of the test.
References
1. Harber P and Lockey JE. Pulmonary function testing in pulmonary
prevention. Occup Med: State of the Art Reviews 6(1): 69-79, 1991.
2. American Thoracic Society: Respiratory Protection Guidelines. Am J Respir
Crit Care Med 154:1153-1165, 1996.
3. Boehlecke BA: Respirators. In: Occupational and Environmental Respiratory
Disease, pp. 963-971, Harber P, Schenker MB, and Balmes JR, editors. Mosby-Year
Book, Inc., St Louis, MO, 1996.
4. U.S. Code of Federal Regulations. Title 29, Part 1910.134 (e), Respiratory
Protection, 1998.
5. USDOD: Occupational Medical Surveillance Manual, Publication No. DOD
6055.5-M, May 1998, available through:
http://www.osha-slc.gov/SLTC/medicalsurveillance
6. USDHHS, PHS, CDC, NIOSH: NIOSH Criteria Documents Plus CD-ROM, Publication
No. PB 97-502-082, 1997.
7. U.S. Code of Federal Regulations. Title 20, Part 404, Subpart P, App. 1,
Federal Old Age, Survivors, and Disability Insurance, Determining Disability and
Blindness, Listing of Impairments, 1985.
8. U.S. Code of Federal Regulations. Title 20, Part 718.103, Federal Coal
Mine Health and Safety Act of 1969, as Amended, Part 718 - Standards for
determining coal miners' total disability or death due to pneumoconiosis, 1983.
9. American Medical Association: Guides to the Evaluation of Permanent
Impairment. Fourth Edition. Chicago: American Medical Association, 1993.
10. American Thoracic Society: Evaluation of Impairment/Disability Secondary
to Respiratory Disorders. Am Rev Resp Dis 133: 1205-1209, 1986.
11. American Thoracic Society: Single-breath carbon monoxide diffusing
capacity (transfer factor): Recommendations for a standard technique - 1995
update.. Am J Respir Crit Care Med 152:285-298, 1995.
12. Morris AH, Kanner RE, Crapo RO, Gardner RM: Clinical pulmonary function
testing: a manual of uniform laboratory testing procedures, 2nd edition.
Intermountain Thoracic Society, Salt Lake City, UT, 1984.
13. Wasserman Karlman et al: Principles of exercise testing and
interpretation, 2nd edition. Lea & Febiger, 1994.
14. Sterk PJ, Fabbri LM, Quanjer PH et al: Airway responsiveness.
Standardized challenge testing with pharmacological, physical, and sensitizing
stimuli in adults. Report of working party standardization of lung function
tests, European Community for Steel and Coal. Official statement of the European
Respiratory Society. Eur Respir J 6 (Suppl 16): 53-83, 1993.
15. Graham WG, et al. Pulmonary function loss in Vermont granite workers. A
long-term follow-up and critical reappraisal. Am Rev Respir Dis 123(1):25-8,
1981.
16. Townsend MC: The effects of leaks in spirometers on measurements of
pulmonary function: the implications for epidemiologic studies. J Occup Med
26(11):835-41, 1984.
17. Townsend MC, Morgan J, Durkin D, et al: Quality control aspects of
pulmonary function testing in the Multiple Risk Factor Intervention Trial. Cont
Clin Tri 7:179S-192S, 1986.
18. Townsend MC: Standardized PFTs in the occupational setting, Parts I and
II. NAOHP Clinical Care Update 4(19-20), Santa Barbara, CA, 1998.
19. Scanlon PD and Hankinson J: Physiologic Methods. In: Occupational and
Environmental Respiratory Disease, pp. 39-54, Harber P, Schenker MB, and Balmes
JR, editors. Mosby-Year Book, Inc., St Louis, MO, 1996.
20. Gardner RM, chairman: ATS statement - Snowbird workshop on
standardization of spirometry. Am Rev Resp Dis 119: 831-838, 1979.
21. American Thoracic Society: Standardization of spirometry - 1987 Update.
Am Rev Resp Dis 136: 1285-1298, 1987.
22. American Thoracic Society. Standardization of spirometry: 1994 Update. Am
J Resp Crit Care Med 152: 1107-1136, 1995.
23. American Thoracic Society: Lung Function Testing: Selection of Reference
Values and Interpretative Strategies. Am Rev Resp Dis 144: 1202-1218, 1991.
24. Hiebert T, Miles J, and Okeson GC: Contaminated aerosol recovery from
pulmonary function testing equipment. Am J Respir Crit Care Med 159: 610-612,
1999.
25. Kendrick AH, Smith EC, and Leeming JP: Bacterial colonization as a
potential source of nosocomial respiratory infections in spirometers. Eur Respir
J 10:1694-1695, 1997.
26. Leeming JP, Pryce-Roberts DM, Kendrick AH, et al: The efficacy of filters
used in respiratory function apparatus. J Hosp Infect 31:205-210, 1995.
27. Rutala DR, Rutala WA, Weber DJ et al: Infection risks associated with
spirometry. Infect Control Hosp Epidemiol 12:89-92, 1991.
28. U.S. Code of Federal Regulations. Title 29, Part 1910.1043 (h) and
Appendix D, Cotton Dust, revised 1985.
29. Horvath EP Jr (Ed.): Manual of Spirometry in Occupational Medicine.
Atlanta: CDC, NIOSH, 1981.
30. Enright, PL and Hyatt, RE. Office Spirometry. Philadelphia: Lea &
Febiger, 1987.
31. Hankinson JL. Instrumentation for spirometry. Occup Med: State of the Art
Reviews 8(2): 397-407, 1993.
32. Ruppel GL. Manual of pulmonary function testing. St. Louis: Mosby-Year
Book Inc., 1998.
33. Hankinson, JL. Beyond the Peak Flow Meter: Newer technologies for
determining and documenting changes in lung function in the workplace. Occup
Med: State of the Art Reviews, 2000, In press.
34. Hankinson, JL and Gardner, RM. Standard waveforms for spirometer testing.
Am Rev Resp Dis 126:362-4, 1982.
35. Nelson SB, Gardner RM, Crapo RO et al: Performance evaluation of
contemporary spirometers. Chest 97:288-297, 1990.
36. Hankinson JL and Wagner GR: Medical screening using periodic spirometry
for detection of chronic lung disease. Occ Med: State of the Art Rev 8(2):
353-361, 1993.
37. Enright PE, Johnson LR, Connett JE, Voelker H, and Buist AS: Spirometry
in the Lung Health Study. 1. Methods and quality control. Am Rev Resp Dis 143:
1215-1223, 1991.
38. Hankinson JL and Bang KM: Acceptability and reproducibility criteria of
the American Thoracic Society as observed in a sample of the general population.
Am Rev Resp Dis 143: 516-521, 1991.
39. Enright PL: Surveillance for lung disease: Quality assurance using
computers and a team approach. Occ Med: State of the Art Rev 7(2): 209-225,
1992.
40. U.S. Code of Federal Regulations. Title 29, Part 1910.1020, Access to
employee exposure and medical records, 1988.
41. The National Lung Health Education Program Executive Committee.
Strategies in Preserving Lung Health and Preventing COPD and Associated
Diseases: The National Lung Health Education Program (NLHEP). Chest 113
(suppl):123S-155S, 1998.
42. Hankinson JL, Castellan RM, Kinsley KB, and Keimig DG: Effect of
Spirometer Temperature on Measurement of FEV1 Shift Changes. JOM
28(12):1222-1225, 1986.
43. National Institute of Occupational Safety and Health, Division of
Respiratory Disease Studies: NIOSH Spirometry Training Guide, 1997.
44. Kellie SE, Attfield MD, Hankinson JL, Castellan RM: The ATS spirometry
variability criteria: Associations with morbidity and mortality in an
occupationally exposed cohort of coal miners. Am J Epidem 125:437-444, 1987.
45. Eisen EA: Standardizing spirometry: Problems and prospects. Occ Med:
State of the Art Rev 2: 213-226, 1987.
46. Pennock BE, Rogers RM, McCaffree DR: Changes in measured spirometric
indices - what is significant? Chest 80:97, 1981.
47. U.S. Code of Federal Regulations. Title 29, Part 1910.1001 (l), Asbestos,
revised 1994.
48. U.S. Code of Federal Regulations. Title 29, Part 1910.1028 (i), Benzene,
1987.
49. American College of Occupational and Environmental Medicine. Guidelines
for Spirometry in the Occupational Setting. JOM 34 (5), 1992.
50. American Association of Occupational Health Nurses. Guidelines for
Implementation of a Respiratory Surveillance Program. October, 1989. Atlanta,
GA.
51. Morris JF, Koski A, Johnson L: Spirometric standards for healthy
nonsmoking adults. Am Rev Resp Dis 103: 57-67, 1971.
52. Crapo RO, Morris AH, Gardner RM: Reference spirometric values using
techniques and equipment that meet ATS recommendations. Am Rev Resp Dis 123:
659-654, 1981.
53. Knudson RJ, Lebowitz MD, Holberg CJ, et al: Changes in the Normal Maximal
Expiratory Flow-Volume Curve with Growth and Aging. Am Rev Resp Dis 127:
725-734, 1983.
54. Ghio AJ, Crapo RO, Elliott CG: Reference equations used to predict
pulmonary function at institutions with respiratory disease training programs in
the United States and Canada: a survey. Chest 63:136-145, 1990.
55. Knudson RJ, Slatin RC, Lebowitz MD, et al: The maximal expiratory
flow-volume curve: Normal standards, variability, and effects of age. Am Rev
Resp Dis 113: 587-600, 1976.
56. Hankinson JL, Odencrantz JR, Fedan KB: Spirometric reference values from
a sample of the general U.S. population. Am J Resp Crit Care Med 159:179-187,
1999.
57. Hankinson JL, Kinsley KB, Wagner GR: Comparison of spirometric reference
values for Caucasian and African American blue-collar workers. JOEM
38(2):137-143, 1996.
58. Glindmeyer HW, Lefante JJ, McCollester C, et al: Blue-collar normative
spirometric values for Caucasian and African-American men and women aged 18 to
65. Am J Resp Crit Care Med 151:412-422, 1995.
59. Coultas DB, Gong H, Grad R, et al: State of the Art: Respiratory diseases
in minorities of the United States. Am J Resp Crit Care Med 149:S93-S131, 1993.
60. Hankinson JL: Pulmonary Function Testing in the Screening of Workers:
Guidelines for Instrumentation, Performance, and Interpretation. JOM 28(10):
1081-1092, 1986.
61. Berry G: Longitudinal observations. Their usefulness and limitations with
special reference to the forced expiratory volume. Bull Physiopathol Respir
(Nancy) 10(5): 643-56, 1974.
62. Burrows B, Lebowitz MD, Camilli AE, Knudson RJ: Longitudinal changes in
forced expiratory volume in one second in adults. Methodologic considerations
and findings in healthy nonsmokers. Am Rev Respir Dis. 133(6): 974-80, 1986.
63. Clement J, Van de Woestijne: Rapidly decreasing forced expiratory volume
in one second or vital capacity and development of chronic airflow obstruction.
Am Rev Respir Dis 125(5): 553-8, 1982.
64. Vollmer WM: Reconciling cross-sectional with longitudinal observations on
annual decline. Occup Med: State of the Art Reviews 8(2): 339-351, 1993.
65. Buist AS, Vollmer WM: The use of lung function tests in identifying
factors that affect lung growth and aging. Stat Med 7(1-2): 11-8, 1988.
66. Camilli AE, Burrows B, Knudson RJ, Lyle SK, et al: Longitudinal changes
in forced expiratory volume in one second in adults. Effects of smoking and
smoking cessation. Am Rev Respir Dis 135(4): 794-9, 1987.
67. Tager IB, Segal MR, Speizer FE, Weiss ST: The natural history of forced
expiratory volumes: Effect of cigarette smoking and respiratory symptoms. Am Rev
Respir Dis 138(4): 837-49, 1988.
68. Ware JH, Dockery DW, Louis TA, Xu X, et al: Longitudinal and
cross-sectional estimates of pulmonary function decline in never-smoking adults.
Am J Epidemiol 132(4): 685-700, 1990.
69. National Fire Protection Association. Medical Requirements for Fire
Fighters, NFPA 1582, 1997 edition. National Fire Protection Association, Quincy,
MA, 1997.
70. USDHHS, PHS, CDC, NIOSH: Criteria for a Recommended Standard:
Occupational Exposure to Respirable Coal Mine Dust. September, 1995.
71. American Thoracic Society: Guidelines for the Evaluation of
Impairment/Disability in Patients with Asthma. Am Rev Resp Dis 147: 1056-1061,
1993.
72. USDHHS, PHS, NIH, NHLBI: National Asthma Education and Prevention Program
(NAEPP) Expert Panel Report 2: Guidelines for the Diagnosis and Management of
Asthma, Clinical Practice Guidelines. 1997, NIH Publication No. 97-4051.
73. USDHHS, PHS, NIH, NHLBI: Global Initiative for Asthma: Global Strategy
for Asthma Management and Prevention: NHLBI/WHO Workshop Report. Bethesda, MD,
1995. NIH Publication No. 95-3659.
74. Buist AS, Connett JE, Miller RD, Kanner RE, et al: Chronic Obstructive
Pulmonary Disease Early Intervention Trial (Lung Health Study): Baseline
Characteristics of Randomized Participants. Chest 103(6): 1863-1872, 1993.
75. Ghio AJ, Castellan RM, Kinsley KB, and Hankinson JL: Changes in Forced
Expiratory Volume in One Second and Peak Expiratory Flow Rate across a Work
Shift among Unexposed Blue Collar Workers. Am Rev Resp Dis 143: 1231-1234, 1991.
76. Burge PS: Use of serial measurements of peak flow in the diagnosis of
occupational asthma. Occupational Medicine: State of the Art Reviews 8(2):
279-294, 1993.
77. American College of Chest Physicians Consensus statement: Assessment of
Asthma in the Workplace. Chest 108:1084-1117, 1995.
78. Gannon PFG and Burge PS: Serial peak expiratory flow measurement in the
diagnosis of occupational asthma. Eur Respir J 10 (Suppl 24): 57s-63s, 1997.
79. Moscato G, et al. statement on self-monitoring of peak expiratory flows
in the investigation of occupational asthma. Subcommittee on Occupational
Allergy of the European Academy of Allergology and Clinical Immunology. American
Academy of Allergy and Clinical Immunology. European Respiratory Society.
American College of Allergy, Asthma and Immunology. Eur Respir J 8(9):1605-10,
1995.
80. European Academy of Allergology and Clinical Immunology Subcommittee on
"Occupational Allergy": Guidelines for the diagnosis of occupational asthma.
Clin Exp Allergy 22:103-108, 1992.
81. Quackenboss JJ, Lebowitz MJ, Krzyzanowski M: The normal range of diurnal
changes in peak expiratory flow rates. Am Rev Resp Dis 143:323-330, 1991.
82. Enright PL, Lebowitz MD, and Cockroft DW: Physiologic Measures: Pulmonary
Function Tests. Asthma Outcome. Am J Respir Crit Care Med 149: S9-S18, 1994.
83. Bernstein DI: Clinical assessment and management of occupational asthma.
In: Bernstein IL, Chang-Yeung M, Malo J-L, et al, editors: Asthma in the
Workplace, pp. 103-123, New York, 1993, Marcel Dekker, Inc.
84. Smith AB, Castellan RM, Lewis D, et al: Guidelines for the epidemiologic
assessment of occupational asthma. Report of the Subcommittee on the
Epidemiologic Assessment of Occupational Asthma, Occupational Lung Disease
Committee. J Allergy Clin Immunol 84: 794-805, 1989.
85. Burge PS: Physiologic Assessment of Occupational Asthma. In: Bernstein
IL, Chang-Yeung M, Malo J-L, et al, editors: Asthma in the Workplace, New York,
1993, Marcel Dekker, Inc.
86. Hankinson JL, Das MK: Frequency response of portable PEF meters. Am J
Resp Crit Care Med 152:702-706, 1995.
Arizona Payday Loan
|